Instruction Manual
IM-106-5500, Original Issue
August 2005
CCO 5500
3-8
NOTE
All operating parameters are entered in the setup mode.
Section 5, Normal Operation, lists all parameters in full. Basic details are 
given here for configuration purposes. 
To aid configuration and to record any subsequent changes to the operating 
parameters, it is recommended that Table 3-1. Instrument Settings, at the end 
of this procedure be completed to provide a record of the instrument setup. 
1. Press the MODE key until the number 5 is displayed in the top left-hand 
corner. 
2. Enter 0000 for the security code - this is the default code set at the 
factory. 
3. After the correct code has been entered the six sub-modes are 
accessed by using the ARROW keys and pressing ENTER when the 
required option is displayed. 
4. First the instrument averages need setting. 
a. Press the ENTER key when this 
display is shown, the display will now 
show one of the averages. Use the 
ARROW keys to select the average time 
that requires setting, and press the ENTER key to access the value. 
Change the value by using the ARROW keys and input by pressing the 
ENTER key. 
b. Set the seconds averaging stack to the required value. This is limited 
to within 10 to 60 seconds in 10-second intervals. 
c. Set the minutes averaging stack to the required value. This is limited 
to within 1 to 60 minutes in 1-minute intervals. 
d. Set the hours averaging stack to the required value. This is limited to 
within 1 to 24 hours in 1-hour intervals. 
e. Set the days averaging stack to the required value. This is limited to 
within 1 to 30 days in 1-day intervals. 
5. The analogue current loop output is set 
up in the next menu. Press the ENTER 
key while this display is shown to select 
it, then press the ARROW keys to step 
through the available options. Press the ENTER key to enter each 
option and change the displayed parameter. 
a. Base of Output 
An origin of 0 or 4mA can be set for 
the current loop output. The ARROW 
keys will 'toggle' between these two 
options. Press the ENTER key to enter 
the new value. 
b. Averaging Time of the Output 
5 SET AVERAGES
5 CONFIGURE O/P
5 CONFIGURE O/P
OUTPUT = 4 to 20 mA