Emerson IM-106-5500 Manual Do Utilizador

Página de 108
Instruction Manual
IM-106-5500, Original Issue
August 2005
4-11
CCO 5500
Figure 4-4. Set Up Mode Tree
Set Averages 
Four separate averages are calculated within the instrument. These are 
defined in units of seconds, minutes, hours and days. Any of the four 
averaging stacks can be used to provide the analog output of the instrument. 
Each averaging time can be set within predefined limits. 
Enter
Press Enter on Exit option
Set Averages
Seconds
Minutes
Hours
Days
10 to 60s @ 10s intervals
01 to 60m @ 01m intervals
01 to 24h @ 01h intervals
01 to 30d @ 01d intervals
Configure O/P
Output
Average
Units
Span
Fault Condition
Set mA O/P
0 or 4-20mA
Take average from secs, mins, hours or days stack
ppm - mg/m3 - mg/Nm3
0 to 9999 span (ppm or mg/m3 - depends on above)
Measured, zero mA, FS (20mA) or fix value
Set up current output levels
Parameters
Security No.
Identity No.
Path Length
Alarm Source
Level
Cal. Factor
Plant Status
4-digit code
01 to 30 (unique number for serial data, I/Ps & O/Ps)
Gas path length in mm
Alarm from secs, mins, hours or days stack
0 to 9999 (ppm or mg/m3 - depends on units)
Determines basic calibration of the instrument
Contact, serial, temp., O2 or Det. level
Normalization
Temperature
Oxygen
Pressure
Water Vapor
Set up normalization I/P & std. levels for temp.
Set up normalization I/P & std. levels for O2
Set up normalization I/P & std. levels for pressure
Set up normalization I/P & std. levels for H2O
Reset Average
Yes
No
Reset the four stack averages
Calibrate
Set Detectors
Span Adjust
Calibrate
Detector D2
Adjust sensitivity
Conduct a basic calibration
Exit options not shown