Emerson IM-106-5500 Manual Do Utilizador

Página de 108
Instruction Manual
IM-106-5500, Original Issue
August 2005
CCO 5500
4-12
1. Press the ENTER key when this display 
is shown; the display will now show one 
of the averages. Use the ARROW keys to 
select the average time that requires 
changing, and press the ENTER key to change it. The value can now 
be changed using the ARROW keys and input by pressing the ENTER 
key. 
2. Set the seconds averaging stack to the 
required value. This is limited to within 10 
to 60 seconds in10-second intervals.
3. Set the minutes averaging stack to the 
required value. This is limited to within 1 
to 60 minutes in 1-minute intervals. 
4. Set the hours averaging stack to the 
required value. This is limited to within 1 
to 24 hours in 1-hour intervals. 
5. Set the days averaging stack to the 
required value. This is limited to within 1 
to 30 days in 1-day intervals. 
Configure O/P 
The analog current loop output is set up from 
this mode. Press the ENTER key while this 
display is shown to select it, then press the 
ARROW keys to step through the available 
options. Press the ENTER key to enter each option and change the displayed 
parameter. 
Base of Output 
An origin of 0 or 4 mA can be set for the current 
loop output. The ARROW keys will 'toggle' 
between these two options. Press the ENTER 
key to enter the new value. 
Averaging Time of the Output 
Any of the four averaging stacks (seconds, 
minutes, hours and days) may be used for the 
analog output. They are selected by the 
ARROW keys and entered using the ENTER 
key. 
Output Units 
The analog output can represent the gas 
concentration in units of mg/m
3
, mg/Nm
3
 or 
ppm. The ARROW keys will 'toggle' between 
these three options. Press the ENTER key to 
enter the new value. 
5 SET AVERAGES
5 SET AVERAGES
secs 60
5 SET AVERAGES
mins 60
5 SET AVERAGES
hours 24
5 SET AVERAGES
days 30
5 CONFIGURE O/P
5 CONFIGURE O/P
OUTPUT = 4 to 20mA
5 CONFIGURE O/P
Average
0.1m
5 CONFIGURE O/P
Units
mg/m3