41 
 
   You now have a terminal connection established with the 4070A. Hit the “?” key. You should see a command 
menu in response.  
   If you do not see this menu, check the following: 
• 
On the 4070A, press Mode then Offset and verify that the baud rate is set to 9600. 
• 
Verify that you are using the correct COM port. 
• 
Check your cabling. 
   If you see the command menu, you have verified that the 4070A is properly cabled to the computer, and that you 
are certain which COM port you are using. 
   NOTE: In general any terminal software capable of sending standard ASCII codes over RS-232 terminal could be 
used to control 4070A remotely, please refer to software provider for specific terminal OS/software support issues.   
6.4 Remote Control Operation 
   Each key on the front panel keypad of the 4070A has an associated ASCII character which, when sent to the 
4070A over the serial port, has the same effect as pressing that key on the keypad.  
   At power-up, the 4070A RS-232 baud rate is recalled from stored configuration 0. The factory default for this is 
9600 baud. The remaining serial port parameters are always 1 start bit, 8 data bits, 1 stop bit, and no parity. The 
serial port on the controlling computer must be set to match these values.  
6.5 Remote Control Programming Rules  
   The following guidelines should be observed when writing control software for the 4070A:  
• 
All ASCII commands are case insensitive, i.e. upper and lower case letters are treated equally by the 
4070A.  
• 
When the 4070A has finished executing a command, it will return a command prompt, which is the 
">"character.  The control software should clear its receive buffer, issue a command, and then look for the 
">" character to determine when the command has been executed. If a long string of commands is sent to 
the 4070A, a separate ">" character will be returned for each command as it is executed. 
• 
All commands require less than 300 mS to execute except for the store or recall commands, which require 
several seconds to complete. 
• 
All white space (carriage returns, linefeeds, tabs, spaces and commas) between commands is ignored. 
Invalid commands (ASCII characters that are not listed in the command menu) are likewise ignored. 
• 
A command does not have to finish execution before another can be sent.  All characters received by the 
4070A are buffered up and executed in the order they were received.  The programmer should allow 
sufficient time for the 4070A to execute each command and for the signal output to settle into the desired 
state. 
• 
If the 4070A is reporting data to the control program, it will place a colon (:) character before the data. This 
makes it easy for the control program to parse a returned string and extract the character(s) needed. Some 
examples of returned values are: 
         
        Example 1
   DTMF Detection Mode detected digits:  
           DTMF: #  
      
        Example 2  Hardware/Software version, serial number (the "V" command):  
 
           BK Precision model: 4070A  
           Software Version: c.2  
           Hardware Version: 1.0  
           S/N: F45E3412AC56  
           PM Checksum: 0017829BB903 
         
   When the operating mode of the 4070A is changed, all parameter values for the previous mode are retained in 
memory as long as power is applied to the unit. For example, if Sinewave mode with an output frequency of 5 MHz 
is changed to sweep mode, the Sinewave mode 5 MHz value will be saved before the mode is changed.  Switching 
back to Sinewave mode will restore the 5 MHz value in the frequency field.  If power is turned off, however, all 
parameters are reset to default values.  To save all parameters for a mode, use the "Store/Recall" command.