B&K 4070A Manual Do Utilizador

Página de 88
 
 
BK Precision 4070A User Manual Rev.2.2 
43 
V - Report hardware and software versions
      
   This command reports the hardware and software version. A hexadecimal serial number, and a program memory 
checksum are also reported by this command: 
                          BK Precision model: 4070A  
           Software Version: c.2  
           Hardware Version: 1.0  
           S/N: F45E3412AC56  
           PM Checksum: 0017829BB903
      
K1.0 - Enable, Disable front panel keys and rotary knob  
   This command is used to disable or enable the front panel keypad and rotary knob.  To disable the front panel and 
knob, issue the two ASCII characters "K0".  To re-enable them, issue the characters "K1".  
   NOTE: On power-up, the keypad, rotary knob, and RS232 port are all enabled and operate together 
simultaneously.  
   If the front panel is not needed during remote operation, it is advisable to disable it with the "K0" command.  This 
would prevent a user from pressing buttons on the front panel of the unit, which could upset a control program by 
unexpectedly changing the state of the 4070A.  
E1,0 - Enable, Disable LCD echo to terminal  
   This command enables (or disables) a feature where the 4070A prints the contents of the LCD display to the 
terminal port whenever the display changes.  It is useful to enable this feature when you cannot see the LCD display.  
It is also used by the point-and-click Remote Control Software for Windows.  
   The command echoes the LCD display by sending two lines of ASCII text, containing all 80 characters of the 
display.  Two lines of 40 ASCII characters are always printed, even if a line on the display is blank. Each line is 
enclosed by the double quote (") character, and each line begins and ends with a carriage return and linefeed 
sequence.  The location of the cursor within the display is also reported as a hex value. 
  A typical LCD display echo in Sinewave mode, for exa mple, would appear on a terminal as follows:  
Cursor: 4B 
"Sinewave mode                           "  
" 1,000,000.00 Hz               -10.0 dBm"  
   The cursor positions for the top line of the display begin with 0 for the upper left corner, and end with hex 27 as 
the last position on in the top right corner.  The cursor positions for the bottom line of the display begin with hex 40 
for the lower left corner, and end with hex 67 as the last position on in the bottom right corner. If the cursor is 
currently off, then the word "off" is printed instead of a hex value. To enable LCD echoing, send the two ASCII 
characters "E1".  To disable LCD echoing, send "E0".   
   NOTE: Each time the "E1" command is issued, the display is re-printed to the terminal. You can use this 
command as a "refresh" function to observe the state of the LCD display at any time. To read the LCD display 
exactly one time, you may issue the command sequence E1E0. This will send the contents of the display exactly one 
time to the serial port and then disable further printing of the display contents.