Service    Appendix C 
Series N5700 User’s Guide 
117 
Symptom Check 
Action 
No output.  
All displays and indicators are blank. 
Is the AC power cord defective? 
 
Is the AC input voltage within range? 
Check continuity. Replace if necessary. 
 
Check AC input voltage. Connect to 
appropriate voltage source. 
 
Output is present momentarily, but shuts 
off quickly. Display indicates AC. 
Does the AC source voltage sag when a load 
is applied? 
Check AC input voltage. Connect to 
appropriate voltage source. 
 
Output is present momentarily, but shuts 
off quickly. Display indicates OUP. 
Is the power supply configured for remote 
sensing? 
Check if the positive or negative load wire 
is loose. 
 
Output voltage will not adjust. 
Front panel CC LED is on. 
Is the power supply in constant current 
mode? 
Check the current limit setting and load 
current. 
 
Output voltage will not adjust. 
Front panel CV LED is on. 
Is the output voltage being adjusted above 
the OVP setting or below the UVL setting? 
Set the OVP or UVL so that they will not 
limit the output. 
 
Output current will not adjust. 
Front panel CV LED is on. 
Is the unit in constant voltage mode? 
Check the current limit and voltage 
setting. 
 
Large ripple present in output. 
Is the power supply in remote sense? 
 
 
Is the voltage drop on the load wire high? 
Check load and sense wires connection 
for noise and impedance effects. 
 
Minimize the drop on the load wires. 
 
No output.  
Display indicates OUP. 
Over-voltage circuit has tripped. 
Turn off the POWER switch. Check load 
connections. If analog programming is 
used, check if the OVP is set lower than 
the output. 
 
No output.  
Front panel PROT indicator is blinking. 
Display indicates EΠA? 
 
 
Display indicates SO? 
 
 
Display indicates O7P? 
 
 
 
Display indicates OCP? 
Check connector J1 ENABLE connection. 
Also check SW1 switch setting. 
 
Check connector J1 Output Shut-Off 
connection. 
 
Check if air intake or exhaust is blocked. 
Check if unit is installed next to heat-
generating equipment. 
 
Check OCP setting and load current. 
 
Poor load regulation. 
Front panel CV LED is on. 
Are sense wires properly connected? 
Connect sense wires according to 
instructions in chapter 2. 
 
Front panel controls are nonfunctional. 
Is the power supply in Local Lockout mode? 
Turn off the POWER switch and wait until 
the display turns off. Turn on the POWER 
switch and press the REM/LOC button.