Appendix C    Service 
118 
Series N5700 User’s Guide  
Error Messages 
Displaying the SCPI error queue 
The entire error queue is read, then emptied, using the following 
command: SYST:ERR? 
Error List 
The following table documents the various error messages that the 
power supply supports: 
 
Error Device-dependent 
Errors 
(these errors set Standard Event Status register bit #3) 
No error 
This is the response to the ERR? query when there are no errors. 
100 
Too many channels 
You have specified more channels than are installed in the mainframe. 
101 
Calibration state is off  
Calibration is not enabled. The instrument will not accept calibration commands. 
102 
Calibration password is incorrect  
The calibration password is incorrect. 
104 
Bad sequence of calibration commands  
Calibration commands have not been entered in the proper sequence. 
114 
CAL:DATE must be yyyy/mm/dd  
The calibration date must be entered in the numeric format yyyy=year, mm=month, dd=date 
203 
Compatibility function not implemented  
The requested compatibility function is not available. 
204 
NVRAM checksum error  
A checksum error has occurred in the instrument’s nonvolatile random access memory. 
205 
NVRAM full  
The nonvolatile random access memory of the instrument is full. 
206 
File not found  
The internal calibration file or the internal channel attribute file was not found in NVRAM. 
209 
Output communications failure  
A hardware failure has occurred on the power supply. 
302 
Option not installed  
The option that is programmed by this command is not installed. 
351 
VOLT setting conflicts with VOLT:PROT setting  
Attempted to program the voltage above the over-voltage protection setting. 
352 
VOLT:PROT setting conflicts with VOLT setting  
Attempted to set the over-voltage protection below the voltage setting. 
353 
VOLT setting conflicts with VOLT:LIM:LOW setting  
Attempted to program the voltage below the under-voltage limit setting. 
354 
VOLT:LIM:LOW setting conflicts with VOLT setting  
Attempted to set the under-voltage limit above the voltage setting