Tektronix TDS460A Manual Do Utilizador
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–51
ACQuire:NUMACq?
ACQuire
:
NUMACq
?
<NR1>
ACQUIRE:NUMACQ?
might return
might return
350, indicating that 350 acquisitions took place since an AC-
QUIRE:STATE RUN command was executed.
ACQuire:NUMAVg
Sets the number of waveform acquisitions that make up an averaged waveform.
This is equivalent to setting the Average count in the Acquisition Mode side
menu.
This is equivalent to setting the Average count in the Acquisition Mode side
menu.
Acquisition
ACQuire:MODe
ACQuire:NUMAVg <NR1>
ACQuire:NUMAVg?
NUMAVg
?
<Space>
<NR1>
:
ACQuire
<NR1> is the number of waveform acquisitions, from 2 to 10,000.
ACQUIRE:NUMAVG 10
specifies that an averaged waveform will show the result of combining 10
separately acquired waveforms.
specifies that an averaged waveform will show the result of combining 10
separately acquired waveforms.
ACQUIRE:NUMAVG?
might return
might return
75, indicating that there are 75 acquisitions specified for averaging.
Syntax
Related Commands
Examples
Group
Related Commands
Syntax
Arguments
Examples