Select channel(s) by 
fitting jumper (s) 
between appropriate 
terminals. (Current 
mode only.) Jumper 
shown fitted to 
‘enable’ Channel 4. 
Jumper shown fitted 
in the unconnected 
‘stored’ 
 
position. 
Figure 4-3.  Location of Jumpers on Circuit Board 
Do not store the jumpers between any other pins in the block 
as this may switch the module into a factory calibration mode, 
which may result in abnormal operation and change of output 
accuracy.  
CAUTION 
The jumper settings can be overridden by the datalogger program, if required, 
using special commands. Please contact Campbell Scientific for details. 
4.4  SDM Address Setting 
Each SDM-CVO4 can have 1 of 16 addresses. The factory-set address is 00. 
Table 4-2 shows switch position and the corresponding address. Figure 4-4 shows 
the position of the switch on the right-hand end of the casing. Note that you will 
have to remove the right-hand mounting bracket to gain access to this switch.