Agilent Technologies 85225F Manual Do Utilizador

Página de 148
104
Installation and User’s Guide
4
Servicing the System
 
To troubleshoot the system
Most system problems are caused by faulty cabling or switch 
configurations. 
Refer to 
 on page 51 and do the 
following:
Check system connections and settings:
1
system connections to the DUT
2
system interconnections 
3
GPIB cabling 
4
GPIB address settings 
If the cabling and switch configurations are verified correct, do the 
following:
Complete the Agilent E8364B PNA Series vector network analyzer 
operator’s check in 
Complete the Agilent 4156C precision semiconductor parameter 
analyzer self- test in 
Complete the Agilent 4284A self- test by cycling the instrument’s line 
power. If errors occur, refer to Appendix B in the precision LCR meter 
operation manual. The operation manual is included in with the 
Agilent 85225F performance modeling system.
If you suspect trouble with the E2050B, see “Chapter 4 
Troubleshooting” in the E5810A LAN/GPIB gateway installation and 
configuration guide.
Complete the Agilent E5250A self- test and leak test found in Chapter 3 
of the low leakage switch mainframe user’s guide. The user’s guide is 
included with the Agilent 85225F performance modeling system.
If a problem with one of the system components is found, refer to the 
troubleshooting and repair information in the individual instrument’s 
product documentation.
For more information, see