RF Subsystem Functional Verification Test
C
Installation and User’s Guide
125
 
Performing the RF Subsystem Functional Verification Test
If the performance of the Agilent E8364B PNA Series vector network 
analyzer is in question, complete the following steps to verify operation.
To perform the RF subsystem functional verification test
Step
Action
Notes
Activate the network 
analyzer.
Connect the network analyzer to 
line power.
Switch on the line power.
Warm-up the network 
analyzer for 1 hour.
Wait 1 hour before continuing to 
step 3.
Run the Operator’s Check.
In the network analyzer System 
menu, point to Service and click 
Operator’s Check.
Click Configure.
On the pull down menu, select 
Automatic.
Click Start-Port 1.
Follow the program prompts.
f
When prompted, place a short on 
Port 1 and click OK.
Check PASS/FAIL status.
Continue to follow the prompts for 
the Attenuator Range tests.
i
Check PASS/FAIL status for each 
attenuator range.
j
Click Start-Port 2.
Repeat actions e through i for Port 2.
l
Click Exit to end the Operator’s 
Check.
This opens the Operator’s Check window.
This selects automatic program execution. To 
enable a prompted test sequence, choose 
Prompted to place a prompt before the 
execution of each range test for each 
attenuator.
When the Port 1 test is complete, the 
Operator’s Check window reappears showing 
the test results for Port 1.
When the Port 2 test is complete, the 
Operator’s Check window reappears showing 
the test results for Port 2.
Test the forward reflection 
mode for channel 1.
Connect the test port cable between 
the PORT 1 and PORT 2 connectors.
Press Preset. (By default, the 
instrument measures channel 1 
forward reflection after instrument 
preset.
Inspect the trace shown on the 
display. It should be similar to the 
trace shown to the right.