Agilent Technologies 85225F Manual Do Utilizador

Página de 148
130
Installation and User’s Guide
D
CV Subsystem Functional Verification Test
 
Understanding the CV Subsystem Functional Verification Test
Use the procedure provided in 
 to manually confirm the functionality of the 
Agilent 4284A precision LCR meter. This procedure does not verify 
performance to specification. This procedure confirms that the precision 
LCR meter is ready for performance verification and/or operation by 
confirming the following hardware functionality:
• memory card read/write test
• LED display test
• LCD display test
• handler I/F test
• scanner I/F EEPROM read/write test
• scanner I/F I/O test
• bias current I/F I/O test
Required Tools
See Chapter 10 of the Agilent 4284A precision LCR meter operation 
manual.