Agilent Technologies 85225F Manual Do Utilizador

Página de 148
Verifying System Functionality
3
Installation and User’s Guide
95
 
Performing the System Functional Verification Test
Complete the following steps to verify system functionality using the 
supplied Agilent 8490D 10 dB fixed RF attenuator as the device under 
test.
To perform the system functional verification test
Step
Action
Notes
Switch on power to the 
system.
Complete the steps listed in 
Proper system function is dependent 
upon the order in which the system 
components are switched on.
Connect the device under 
test to the bias networks.
Refer to the following figure.
Connect one end of the 2.4 mm 
female-to-female adapter to the semi-rigid 
cable attached to the RF/DC OUT 
connector of the port 1 bias network. 
Connect male end of the attenuator to the 
other end of the 2.4 mm female-to-female 
adapter.
Connect the female end of the attenuator to 
the semi-rigid cable attached to the RF/DC 
OUT connector of the port 2 bias network.
The 2.4 mm female-to-female adapter 
is a 11900B. Use the 11900B provided 
in the calibration kit.
The port 1 bias network is a 11612V 
K11.
The port 2 bias network is a 11612V 
K21.