Chapter 11
11-2
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Diagnostics
Entering Diagnostic 
Mode
To enter Diagnostic Mode, hold down the ALARM RESET and 100% O
2
 keys 
for approximately 5 seconds while you turn on the ventilator. A message 
appears on the ventilator prompting you to confirm that the patient is 
disconnected before entering Diagnostic Mode (Figure 11-1). Press OK to 
enter Diagnostic Mode.
Figure 11-1: Entering Diagnostic Mode
Once you’ve entered Diagnostic Mode, you can select any of diagnostic 
functions by pressing its button. Change values in Diagnostic Mode the same 
way you change ventilator settings.