Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Diagnostics
Extended Self Test 
(EST)
EST verifies the overall functional integrity of the ventilator by testing all 
critical hardware subsystems and components. Perform EST between patients 
as part of preventive maintenance, a performance verification, or if the 
operational integrity of the ventilator is in question. EST is typically run by 
qualified trained personnel.
Equipment required to run EST:
Patient circuit
Plug for the patient wye
High pressure O
2
 source
Follow these steps to run EST:
1. Enter Diagnostic Mode as described “Entering Diagnostic Mode” on 
2. Connect a patient circuit to the ventilator.
3. Press the EST button on the diagnostic screen (Figure 11-1).
4. Press START EST to begin EST.
5. Follow the prompts.
6. When EST completes successfully, press the OK button.
Hardware
The hardware function allows a trained service technician to operate critical 
components separately to help identify a faulty component in the event that 
the ventilator fails SST, EST, or performance verification testing. For more 
information on the Hardware function, see the Esprit/V200 Ventilator Service 
Manual, P/N 580-1000-02.
Software
The software function displays the ventilators serial number, software part 
numbers, and version numbers and the part numbers of other critical 
components.