Chapter 12
Technical Specifications
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-5
 
Patient Data Screen
Monitored patient data is displayed when Patient Data is pressed. Patient data 
ranges, resolution, units, and accuracy specifications are provided in the table 
below. Minute volume data is based on an eight breath average.
Non-Invasive Ventilation
Value Entry Reasonability Checks
Item
Limit Exceeded
Message
Changed
Apnea Rate
I:E Ratio>4:1
I:E Ratio must be less than 
or equal to 4:1. Check I-
Time.
Readjust the Apnea Rate or 
adjust I-Time or Rate.
I-Time
I:E Ratio>4:1
I:E Ratio must be less than 
or equal to 4:1. Check 
Apnea Rate.
Readjust the I-Time or 
adjust Apnea Rate or Rate.
IPAP
IPAP<EPAP
IPAP must be greater than 
or equal to EPAP.
Readjust IPAP higher or 
adjust EPAP lower.
EPAP
EPAP>IPAP
EPAP must be less than or 
equal to IPAP.
Readjust EPAP lower or 
adjust IPAP higher.
Rise Time
Rise Time>I Time
Rise Time must be less 
than I-Time.
Decrease the Rise Time or 
increase the I-Time.
Table 12-8: Non-Invasive Ventilation Value Entry Reasonability Checks
General Value Entry Reasonability Checks
Item
Limit Exceeded
Message
Changed
Apnea
Apnea Rate<Set Rate Apnea Rate must be 
greater than or equal to the 
Set Rate.
Increase the Apnea Rate or 
decrease the Rate.
Table 12-9: General Value Entry Reasonability Checks
Patient Data Range, Resolution, Units & Accuracy
Patient Data
Display Range
Units
Resolution
Accuracy
Exhaled Minute
Volume
0.00 to 99.9
L
0.01 for 0.00-9.99;
0.1 for 10.0-99.9
±10%
Exhaled Tidal Volume
0 to 9999
mL
1
±10%
Spontaneous Minute 
Volume
0.00 to 99.9
L
0.01 for 0.00-9.99;
0.1 for 10.0-99.9
±10%
Rapid Shallow Breathing 
Index
0 to 500
Bpm/L
1
±10%
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 1 of 2)