Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Technical Specifications
Front Panel Keys
I:E
9.9:1 to 1:99
none
0.1 for 9.9:1-1:9.9;
1 for 1:10-1:99
±10%
Peak Inspiration
Pressure
-20.0 to 130
cmH
2
O
(hPa)
1 for 100-130;
0.1 for -20.0 to 99.9
±10%
End Inspiration
Pressure (Pi End)
-20.0 to 130
cmH
2
O
(hPa)
1 for 100-130;
0.1 for -20.0 to 99.9
±10%
Mean Airway
Pressure
-20.0 to 120
cmH
2
O
(hPa)
1 for 100-120;
0.1 for -20.0 to 99.9
±10%
Delivered O
2
%
0.0 to 110
--
1 for 100-110;
0.1 for -20.0 to 99.9
±3 vol%
Total Respiratory Rate
0.0 to 150
Bpm
1 for 10-150;
0.1 for 0.0 to 9.9
±10%
Spontaneous
Respiratory Rate
0.0 to 150
Bpm
1 for 10-150;
0.1 for 0.0 to 9.9
±10%
End Exhalation
Pressure (Pe End)
-20.0 to 99.9
cmH
2
O
(hPa)
0.1
±10%
Patient Leak (Pt Leak)
0.0 to 140
LPM
0.1 for 0.0-99.9
1 for 100-140
Not 
specified
a
Percent Patient Trigger 
(%Pt Trigger)
0.0 to 99.9
%
0.1 for 0.0-99.9
1 for 100
±10%
a. Due to the variable nature of patient leaks, this parameter is an estimate only.
Patient Data Range, Resolution, Units & Accuracy (Continued)
Patient Data
Display Range
Units
Resolution
Accuracy
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)
Front Panel Keys
Alarm Silence
Screen Lock
Alarm Reset
Expiratory Hold
100% O
2
Accept
Manual Breath
Table 12-11: Front Panel Keys