Troubleshooting 67
1530 CLEAR
SCREEN
1540
PRINT "This program should ONLY be completed if your power supply”
1550
PRINT "EEPROM has been replaced or a component that will effect"
1560
PRINT "the calibration AND the alignment of voltage, overvoltage"
1570
PRINT "and current is complete AND unit has passed the performance"
1580
PRINT "test. Enter C to continue, any other key to abort.”
1590 INPUT
Cont_prog$
1600 IF (UPC$(Cont_prog$))< >"C" THEN GOTO Cal_off
1610
1610
!
1620
CLEAR SCREEN
1630 PRINT "Transferring calibration data to factory preset locations."
1640
1640
!
1650 Fact_cal_sour: ! Address of factory calibration data source
1660
1660
DATA 2,6,68,72,20,24,76,80,150
1670
!
1680 Fact_cal_dest : ! Address of factory calibration data destination
1690
1690
DATA 84,88,92,96,100,104,108,112,116
1700
!
1710 Fact_cal_len: ! Length of factory calibration data
1720
1720
DATA 4,4,4,4,4,4,4,4,1
1730
!
1740 RESTORE
Fact_cal_sour
1750
FOR I=1 TO 9
1760
READ Cal_sour_addr(I)
1770
NEXT I
1780
!
1790 RESTORE
Fact_cal_dest
1800
FOR I=1 T0 9
1810
READ Cal_dest_addr(I)
1820 NEXT
I
1830
!
1840 RESTORE
Fact_cal_len
1850
FOR I=1 T0 9
1860
READ Cal_length(I)
1870 NEXT
I
1880
!
1890
FOR I=1 T0 9
! Locations of good data
1900
OUTPUT @Ps;"DIAG:EEPR? ";Cal_sour_addr(I);",";Cal_length(I) ! Read good data
1910
ENTER @Ps;Cal_data$
! Enter good data
1920
OUTPUT @Ps;"DIAG:EEPR";Cal_dest_addr(I);",";Cal_length(I);”,”;Cal_data$
! Write good data to factory preset locations
1930
1930
NEXT I
1940
!
1950
!
1960 Cal_off
1970
1970
CLEAR SCREEN
1980
OUTPUT @Ps;"CaL:STATE OFF"
! Turn off cal mode
1990
!
2000
GOSUB Ps_error
! Check for errors
Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 4 of 5)