Agilent Technologies Model 6652A: 3201A-00381 Manual Do Utilizador

Página de 133
General Information 19
 
 Output Characteristic
 
 General
 
 The power supply can operate in either CV (constant voltage) or CC (constant current) over its voltage and current ratings
(see Table 1-l). The operating locus is shown by the Output Characteristic Curve in Table 1-2. The operating point is
determined by the voltage setting (V
s
), the current setting (I
s
), and the load impedance. Two operating points are shown.
Point 1 is defined by the load line cutting the operating locus in the constant-voltage region. This region defines the CV
mode. Point 2 is defined by the load line cutting the operating locus in the constant-current region. This region defines the
CC mode.
 
 Downprogramming
 
 The power supply can sink current for more rapid down programming in the CV mode. For Series 664xA and 665xA
supplies, this capability is defined by the second quadrant area (-I
s
) of the Output Characteristic Curve. These supplies can
sink about 20% of their maximum rated positive output current. For Series 667xA and 668xA power supplies, this is an
uncharacterized current-sinking area that provides a limited downprogramming capability.
 
 
 Specifications and Supplemental Characteristics
 
 Tables 1-1 through 1-4 list the specifications and supplemental characteristics for the Series 664xA, 665xA, 667xA, and
668xA power supplies. The organization is as follows:
 
 Series                                               Specifications                      Characteristics
 6641A-6645A
Table l-la
Table l-lb
 6651A-6655A
Table 1-2a
Table 1-2b
 6671A-6675A
Table 1-3a
Table 1-3b
 6680A-6684A
Table 1-4a
Table 1-4b
 
 Specifications are performance parameters warranted over the specified temperature range.
 
 Supplemental Characteristics are not warranted but are descriptions of performance determined either by design or type
testing.