Agilent Technologies Model 6652A: 3201A-00381 Manual Do Utilizador

Página de 133
General Information 21
 Table 1-1b. Supplemental Characteristics for Series 664xA
1
 
 
Parameter
 Agilent Model Number
 
 6641A
 6642A
 6643A
 6644A
 6645A
 Output Programming Range (maximum programmable values)
 
   Voltage:
 8.190  V
 20.475  V
 35.831  V
 61.425  V
 122.85  V
   Current:
 20.475  A
 10.237  A
 6.142  A
 3.583  A
 1.535  A
   Overvoltage Protection (OVP):
 
 8.8  V
 22.0  V
 38.5  V
 66.0  V
 132.0  V
 Average  Resolution
 
 
 
 
 
   Voltage:
 2 mV
 5 mV
 10 mV
 15 mV
 30 mV
   Current:
 6  mA
 3  mA
 2  mA
 1.2  mA
 0.5  mA
   Overvoltage Protection (OVP):
 
 13  mV
 30  mV
 54  mV
 93  mV
 190  mV
 Accuracy
 
 
 
 
 
   Overvoltage Protection (OVP):
 160  mV
 400  mV
 700  mV
 1.2  V
 2.4  V
   Analog Programming (VP):*
 
 
 
 
 
                                                0.36% +
 6 mV
 15 mV
 27 mV
 45 mV
 90 mV
   Analog Programming (IP):*
 
 
 
 
 
                                                  7.6% +
 18 mA
 9.2 mA
 
 
 
                                                  1.5% +
 
 
 5.5 mA
 3.2 mA
 1.4 mA
 Current Monitor (+IM):*
 
 
 
 
 
                                                  7.7% +
 65 mA
 32 mA
 
 
 
                                                 1.6 % +
 
 
 8.1 mA
 7.1 mA
 1.8 mA
 *Referenced to supply output
 
 
 
 
 
 Drift Temperature Stability (following a 30-minute warmup, change in output over 8 hours under constant line, load,
and ambient temperature)
 
  Voltage:                                 0.02% +
 0.4  mV
 1  mV
 2  mV
 3  mV
 6  mV
   Current:                                0.02% +
 16 mA
 6 mA
 3 mA
 2 mA
 1 mA
 Temperature Coefficients (change per °C)
 
 
 
 
 
   Voltage:
 60 ppm +
 0.1 mV
 0.2 mV
 0.3 mV
 0.5 mV
 1.1 mV
   +Current:
 95 ppm +
 0.82 mA
 0.41 mA
 0.18 mA
 0.12 mA
 0.04 mA
  Voltage Readback:
 60 ppm +
 0.2 mV
 0.5 mV
 0.75 mV
 1.3 mV
 2.6 mV
  +Current Readback:
 95 ppm +
 1.2 mA
 0.62 mA
 0.33 mA
 0.20 mA
 0.08 mA
  --Current Readback:
 110 ppm +
 1.2 mA
 0.62 mA
 0.33 mA
 0.20 mA
 0.08 mA
  Overvoltage Protection (OVP):
 
 
 
 
 
 
 200 ppm +
 1.6 mV
 3.3 mV
 5 mV
 13 mV
 24 mV
  Analog Programming (VP):
 
 
 
 
 
 
 60 ppm +
 0.1 mV
 0.25 mV
 0.4 mV
 0.7 mV
 1.25 mV
  Analog Programming (IP):
 
 
 
 
 
 
 90 ppm +
 0.56 mA
 0.28 mA
 0.17 mA
 0.1 mA
 0.04 mA
  Current Monitor (+IM):
 
 
 
 
 
 
 75 ppm +
 0.61 mA
 0.3 mA
 0.06 mA
 0.06 mA
 0.02 mA
  Maximum Input Power:
 480 VA; 400 W, 60 W with no load
 
 Notes:      
1
For Performance Specifications, see Table 1-la.
 
 :