Philips M1351A Manual Do Utilizador

Página de 189
Safety Testing
100
Chapter 9    Preventive Maintenance, Care and Cleaning    
n
The Toco display shows 0.
n
TOCO int
, time, date and paper speed are printed on the paper.
3. Gently apply pressure to the syringe plunger. Check that the value on the display and 
paper shows this change in pressure.
If the test fails, repeat using another transducer. If it still fails, refer to 
Chapter 8, “Troubleshooting Flowcharts.”
Note— The Intrauterine Pressure Values can be between -99 and 127 units.
Safety Testing
This section defines the test and inspection procedures applicable to the Series 50 A and 
Series 50 IP fetal monitors. Use the tables in the following section to determine what test 
and inspection results must be reported after an installation or repair has been carried out. 
n
Test Blocks in Table 9-2 tells you when to carry out the safety tests 
n
Test and Inspection Matrix in Table 9-3 tells you how to carry out the safety tests.
Warning
Apply the safety test limits set by the local standards and statutes applicable to the 
country of installation, such as IEC 60601-1, UL2601-1, CAN/CSA-C22.2 No. 601.1-
M90. 
The safety tests defined in this chapter are derived from local and international 
standards but may not be equivalent. They are NOT a substitute for local safety 
testing.
Caution
Successful completion of the patient safety tests does not ensure the correct 
and accurate functioning of the equipment.
Safety Test 
Procedures
The test procedures outlined in this section are to be used only for verifying the safe 
installation or service of the product. “Safety checks at installation refer to safety aspects 
directly related to the installation and setup activities and not to intrinsic safety features that 
have already been checked during final acceptance testing at the factory.” 
[FDA 97-4179 Medical Quality Systems Manual: A Small Entity Compliance Guide (QSR 
manual). First Edition December 1996. Chapter 14: Storage, Distribution and Installation. 
This document may be viewed at http://www.fda.gov].
When running safety tests on the peripheral components of the system, refer to the