Safety Testing
102
Chapter 9    Preventive Maintenance, Care and Cleaning    
When to 
Perform Safety 
Tests
 
Table 9-2M1351A and M1353A: When to perform safety test blocks
Service Event 
Test Block(s) Required
Installation
The product is customer installed.
You will find installation instructions in this 
Installations and Service Guide.
Perform Visual, Power On and Performance 
test blocks (See Table 9-3).
Preventive Maintenance
Preventive Maintenance is the responsibility of 
the customer
For preventive maintenance refer to page 96.
Perform Visual test block (see Table 9-3).
Repair
This Installation and Service Guide contains 
repair instructions for the 50 A and 50 IP moni-
tors.
Perform Visual, Power On and Performance 
test blocks (see Table 9-3)
when power switch is replaced perform S2-Safety 
test block (see page 104), 
when transformer is replaced perform S3-Safety 
test block (see Table 9-3).
When Frontend-Board is replaced perform S3-
Safety test block (see Table 9-3).
Upgrade
This Installation and Service Guide contains 
upgrade instructions for the 50 A and 50 IP 
monitors.
Perform Visual, Power On, Performance and 
Safety test blocks (see Table 9-3).
All other Service Events
Perform Visual, Power On and Performance 
test blocks (see Table 9-3).