Safety Testing
Chapter 9    Preventive Maintenance, Care and Cleaning    
105
Instrument Safety
Test
You must perform the instrument safety test every time you exchange, repair, upgrade or in 
any other way work on the front end board, the power supply, or the power inlet. If you 
intend to connect the monitor to an obstetrical information management system such as 
OB TraceVue, you must perform the instrument safety test with the monitor as a 
standalone unit, before reconnecting it to the system. 
The instrument safety test is made up of four separate tests (see page 104):
n
Protective Earth Test
n
Enclosure Leakage Current - Normal Condition
n
Enclosure Leakage Current - Single Fault Condition
n
Patient Leakage Current - Single Fault Condition
System Safety Test
Whenever you connect a monitor to an obstetrical system such as OB TraceVue you must 
perform the System Test. The System Test consists of the Enclosure Leakage Current tests 
(normal and single fault condition) as described on page 104. You must run this test when 
connecting either a new or repaired monitor.