Safety Testing
Chapter 9    Preventive Maintenance, Care and Cleaning    
103
Test and 
Inspection 
Matrix
Key to Table: P = Pass and F = Fail, X = test result value to be recorded.
Table 9-3 M1351A/ M1353A Fetal Monitor - Test and Inspection Matrix
Test Block 
Name
Test or “Inspection” to Perform
Expected Test 
Results
What to Record 
on Service 
Record (Philips 
Personnel only)
Visual
Inspect the unit, transducers and cables for any 
damage.
Are they free of damage?
If Yes, Visual test is 
passed.
V:P or V:F 
Power On
Power on the unit.
Does the self-test complete successfully?
If Yes, Power On test is 
passed.
PO:P or PO:F 
Performance
Perform the quick test and parameter test!
Do these tests complete without errors?
If Yes, Performance 
Test is passed.
P:P or P:F
Safety(2)
Protective Earth.
See -104 - Safety (2)
With mains cable:
Maximum impedance 
= x (<= 200 mOhms)
S2:P/x or
S2:F/x
Safety(3)
Enclosure Leakage Current - Normal Condition.
See -104 - Safety (4)
Maximum leakage cur-
rent = x1
(<= 100
µ
A)
S2:P/x4 or
S2:F/x4 
Enclosure Leakage Current - S.F.C Open Supply.
See -104 - Safety (5)
Maximum leakage cur-
rent = x2
(<= 500
µ
A)
(Note: maximum leak-
age current in the US = 
300 
µ
A)
S2:P/x2 or
S2:F/x2
Enclosure Leakage Current - S.F.C Open Earth.
See -104 - Safety (6)
Maximum leakage cur-
rent = x3
(<= 500
µ
A)
(Note: maximum leak-
age current in the US = 
300 
µ
A)
S2:P/x3 or
S2:F/x3
Safety S (3)
M1353A only:
ONLY TOCO-input tested, Metron Testconn. 
required. (Metron order # 19528 and Philips 
Adapter M1940A)
Patient Leakage Current - S.F.C. Mains on 
Applied Part.
See -104 - Safety (3)
Maximum Leakage 
current = x
(<=50
µ
A @ 250V or 
<= 20
µ
A @ 120V)
S3(S):P/x or 
S3(S):F/x