Philips M1351A Manual Do Utilizador

Página de 189
Connecting Peripheral Devices
108
Chapter 10    Peripherals    
Connecting Peripheral Devices
If the monitor has the Combined Interface Module, you can connect:
                                                  
l
OB TraceVue
l
Telemetry System
l
and one of the following:
n
Barcode Reader
n
Maternal NIBP Monitor
n
FSpO
2
 Monitor
To connect the peripheral devices:
1. Lift out the connector cover (A).
2. Connect the peripheral devices to the sockets on the Interface Module.
3. Tighten the screws on the connectors.
4. Feed the cables through the hollow in the center of the Interface Module (B) and then 
push the cable clamp into place.
Figure 10-2Connecting Peripheral Devices
Key to diagram:
C
RS232 Serial Interface (Barcode or NIBP or FSPO
2
 Monitor)
D
Telemetry Interface
E
System Interface (M1353-66531E is required for OB TraceVue
Note— To connect a fetal monitor to the obstetrical surveillance system OB 
TraceVue M1383A/B/C, you will need the Combined Interface Module 
M1353-66531 rev. E (part number M1353-66531E) and fetal monitor 
firmware rev. A.02.02 or later. You will also need the Interface Cable 
M1380-61613 (or the equivalent OB TraceVue cable, Option #K13).