RS232 Serial Interface
112
Chapter 10    Peripherals    
To connect a COLIN Model BP-8800 Monitor to the M1351A or M1353A:
1. Use the interface cable supplied by the COLIN Corporation with the NIBP Monitor.
2. Connect the 9-pin end of the cable to the fetal monitor.
3. Connect the 15-pin end of the cable to the COLIN.
4. Secure the interface cable by the screws at the cable connectors.
Figure 10-6COLIN Interface Cable
Baudrates
The baudrates for each external device must be set as follows:
COLIN BP-8800
Baudrate:
4800
Startbit:
1                                 Parity:             Even
Databits:
8
Stopbits:
1
Cable:
Supplied by the COLIN Corporation.
Dinamap
 1846/
8100
Baudrate:
600
Startbit:
1                                 Parity:              None
Databits:
8
Stopbits:
1
Cable:
M1350-61609 or M1350-61608.
Maternal 
Measurements 
on the FHR 
Trace
Each time an external device takes the maternal NIBP measurement it is printed on the 
trace. When you monitor NIBP, an MHR measurement is printed at the same time as the 
NIBP measurement.
         
 
If you set the NIBP Monitor to automatic mode, you must leave a minimum time interval 
between each measurement to print every value. This time interval depends upon the paper 
speed setting.