Philips M1351A Manual Do Utilizador

Página de 189
Configuring the Monitor Using a PC
30
Chapter 4    Configuring the Monitor    
Installing the 
Service 
Program
Before you install the program, ensure that the serial port COM1 is set up as follows: 
       
Transmission Rate
9600 BPS
Parity
None
Word length (bits)
8
Stop-bits
1
This is done by giving the mode command (
mode com1:96,N,8,1
) at the DOS 
prompt. Refer to your MS-DOS Manual for details.
To install the program:
1. Switch on the PC.
2. Create a directory for the program files. For example, to create a directory called 
service
, at the DOS prompt type in:
md service
and press 
Enter
.
3. Insert the program disk into drive A.
4. At the DOS prompt type 
A:
 and press 
Enter
.
5. At the 
A:
 prompt, type:
copy pegserv.exe c:\service
and press 
Enter
. Where 
service
 is the directory you created to contain the 
program files.
The program is copied to 
c:\service
Make sure that the serial interface configuration on the PC is set up as follows: 
Datacom-Configuration
Parameter
Serial
Transmission Rate (BPS) 9600
Word Length (bits)
8
Stop-Bits
1
Parity
None
%ON /%OFF Pacing 
Off 
CTS Line
Regard
DSR Line
Regard
DCD Line
Ignore
Power to Interface
Off