Philips M1351A Manual Do Utilizador

Página de 189
Tests and Error Messages
66
Chapter 7    Tests and Error Messages    
n
Err xxx
 
, time and date are printed on the paper after 13 seconds, and 
then every 10 minutes. 
(
xxx
 is the number of the error message.)
Switch the monitor off and then on, and if the error occurs again refer to 
Chapter 8, “Troubleshooting Flowcharts.”
 (If the recorder is not on when the 
monitor is switched on, 
Err xxx
 
, time and date are printed when it is 
switched on subsequently.)
l
If a fatal error occurs (for example, if a board is defective): 
n
An error message is displayed for approximately 13 seconds.
n
After 13 seconds, the monitor will try to restart.
Quick Test
This tests the display, recorder and monitor hardware and takes approximately 15 seconds. 
To carry out the test:
         
1. Switch on the monitor and the recorder.
2. Disconnect any transducers from the monitor (and if appropriate, switch off and 
disconnect the telemetry receiver).
3. Press 
Test
 once.
Figure 7-1 System Test Controls
Check that:
n
The displays flash alternately with the indicators and lamps for approximately 13 
seconds.
n
The 
Err xxx
 on/off light blinks in time with the display.
n
A test pattern is printed on the paper.