Chapter 7    Tests and Error Messages    
Parameter Test
This tests the signal to and from the transducer, but not the transducer itself. To carry out 
the test:
           
1. Switch on the monitor and the recorder.
2. Connect the appropriate transducer to each socket.
3. Press and hold 
Test
.
Figure 7-3 Parameter Test Controls
The correct monitor response for each signal is given in the following table. If your response 
is significantly different, refer to Chapter 8, “Troubleshooting Flowcharts.”.
 
If an error occurs:
l
An error message is displayed for ten seconds.
l
Err xxx
 
, time and date are printed on the paper after ten seconds, and then 
every ten minutes.
(
xxx
 is the number of the error message.)
To stop the error annotation printing, switch the monitor off and then on.
Table 7-1
Signal Correct 
Monitor 
Response
US/US1 
190
 is displayed and printed, the Signal Quality Indicator is green, and the 
fetal heartbeat is heard from the loudspeaker.
Toco 
A signal alternating between 
10
 and 
60
 is displayed and printed.
DECG 
200
 is displayed and printed, the Signal Quality Indicator is green, and the 
fetal heartbeat is heard from the loudspeaker.
MECG 
120
 is printed.
US2 
170
 is displayed and printed, the Signal Quality Indicator is green, and the 
fetal heartbeat is heard from the loudspeaker.