C
ALIBRATION
43
8 Check the settings and click Measure.
The Information dialog box appears.
9 Place the ES-1000 in the calibration cradle.
When you place the ES-1000 in the calibration cradle, white point calibration is used to 
calibrate the spectrophotometer and compensate for gradual drifts in the instrument. Make 
sure the sample aperture is in full contact with the white tile on the calibration cradle. If you 
do not place it correctly in the calibration cradle, the spectrophotometer does not return 
accurate measurements.
10 Click OK.
The Measurements dialog box appears. Instructions for measuring the strips appear in the 
Status field.
11 For a more accurate measurement, place several sheets of plain white paper beneath the 
measurement page.
The extra sheets block underlying colors from being read by the instrument.
12 Orient the measurement page so that strips are horizontal and the scan direction (indicated 
by the arrows at the beginning of each strip) is left to right. 
13 Hold the ES-1000 with its length perpendicular to the scan direction, and place the sample 
aperture in the white space at the start of the specified color. 
14 Press and hold the Measure button and wait for a beep or wait for the background color in the 
dialog to change from green to white. 
N
OTE
To hear a beep with a Windows compatible computer, you must install a sound card 
and speaker(s). You may have to turn up the volume on your computer to hear the beep.