4 Basic Operations
23
Remove the objective lens cap and turn the retinal 
camera's power switch to the ON position. 
The POWER lamp lights and the SET lamp blinks. 
The CR-2 AF automatically starts an initialization sequence 
(internal check and initialization of driving parts). After the 
initialization, the parts of the main unit stop in their center 
positions. 
CAUTION
The main unit moves to the center position when the CR-2 AF is turned on. Do not touch the main 
unit while it is moving. 
Power saving function 
If the power is on and the buttons on the operation panel are not operated for over 10 minutes, the 
power saving function is activated. The POWER lamp blinks while in power saving mode. 
To return to normal mode, press any button on the operation panel or move the stage right or left. 
Press and hold the unlock button and slide the stage 
lock to the UNLOCK side. 
Align the mark on the stage lock and the mark on the left side 
of the base. 
The stage lock is released. 
Move the stage toward you with the operation lever. 
Disinfect the forehead rest and replace the chin rest paper. 
If chin rest paper is not being used, disinfect the chin rest also. 
 CAUTION
•  To prevent the risk of infection, wipe the forehead rest with disinfectant ethanol for each patient. 
For details on how to disinfect, consult a specialist.
•  To ensure cleanliness, replace the chin rest paper for each patient.