15.
 
SERVICE PARAMETERS 
15.1.
 
ACCESS TO PARAMETERS 
APPROVED SCALES 
Access  to  the  scale  parameters  and  calibration  is  controlled  in  all  approved  scales 
either  by  limiting  access  to  be  after  the  Calibration  Jumper  is  put  on  the  PCB, 
location  J1,  pins  1  &  2.    In  this  case  the  display  will  show  the  passcode  request 
screen, “ P - - - - “ .  To continue enter a passcode as described below.  
Or  if  the  Calibration  and  Parameters  have  been  enabled  the  user  must  enter  the 
correct password to have access.  See Section 6.0.   
Entering  passcode  0000  will  allow  calibration  as  shown  in  15.1,  entering  1000  will 
allow access to a limited set of parameters as described in section 6.2.
 
NON-APPROVED SCALES 
Non EC Approved scales will allow entry to the parameters if the Tare key is pressed 
during the power on cycle.   The passwords work as above. 
 15.1  USING “0000” TO ENTER THE CALIBRATION PARAMETER 
“Pn” 
 
 
When “Pn” is displayed. 
Enter “0 0 0 0” and press [Tare] 
“UnLoAd ” 
   
Empty  the  platform  by  removing  the  load,  if 
there is any and press [Tare] 
“LoAd” “6” “KiLoS” 
Load the requested calibration weight and press 
[Tare] 
“SPAn”  “PASS” 
If  Calibration  is  complete,  “SPAn  PASS”  will  be