Adam Equipment gbk 6m Manual Do Utilizador

Página de 68
 
52
 | 
P a g e                  
© Adam Equipment Company 2010
 
displayed. Remove the calibration weight. 
Or, “SPAn”  “FAiLEd ” 
This  means  calibration  has  failed.  Remove  the 
calibration weight and repeat the process. 
“JP On” 
Remove  the  jumper  or  shorting  of  the  pins 
whichever  is  used.  The  scale  will  return  to 
normal weighing. 
 
15.2.
 
USING “1000” TO ENTER THE SERVICE PARAMETERS 
Non-Approved indicators: 
  For  the  non  approved  indicator  press  the  [Tare]  key  during  the  display  counting 
when turned on,   
Approved Indicators:  For the Approved version a jumper can  be installed to allow 
access or the Calibration and Parameter Counters must be enabled (see 15.2.10).