50
8.9 
P8.0: Selection of cell constant 
Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.  
Use a cell of K = 1.0 for midrange measurements 
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).  
Use a cell of K = 0.1 for low range measurements (below 20 
µS or 10 ppm).  
The cell included with your meter has a cell constant of K = 
1.0.  
From measurement mode 
1.  Press SETUP key to enter Set Up menu.  
2.  Press the MI/▲ or MR/▼ keys until you view 
parameter P8.0.  
3.  Press the HOLD/ENTER key to enter parameter 
P8.0.  
4.  Press the MI/▲ or MR/▼ keys to select the cell 
constant between K = 1.0, 0.1, or 10.  
5.  Press the HOLD/ENTER key to confirm selection 
and to return to the subgroup menu.  
6.  Press the CAL/MEAS key to return to measurement 
mode.  
 
 
 
cel
1.0
SETUP
SETUP
p 8.0
p 8.0
p 8.0
10.0
SETUP
MI
MR
HOLD
ENTER
K = 
K = 
Figure 43: P8.0-