CON 11/110 
 
51
  
8.10  P9.0: Automatic and Single point Calibration 
You may select an automatic or manual calibration depending on the calibration 
standards that you have. Program 9.0 lets you have the option to do the above as 
well as the option of having a single or a multi point calibration. 
8.10.1 P9.1: Selection of Automatic or Manual Calibration 
The automatic calibration feature lets you calibrate the meter automatically by using 
the appropriate known calibration standard solutions whereas in the manual 
calibration feature non-standard calibration values can be used for calibration.  
NOTE: Automatic calibration is applicable in conductivity 
measurement mode only whereas manual calibration is 
applicable in both conductivity and TDS mode. 
Only from the conductivity measurement mode, 
1.  Press SETUP key to enter Set Up menu.  
2.  Press the MI/▲ or MR/▼ keys until you view 
parameter P9.0.  
3.  Press the HOLD/ENTER key to enter parameter 
P9.1. The upper display will show “ACA”. 
4.  Press the HOLD/ENTER key to enter the automatic 
or manual calibration selection mode. 
5.  Press the MI/▲ or MR/▼ keys to select “YES” or 
“NO”. 
• 
YES activates the automatic calibration. 
• 
NO activates the manual calibration. 
6.  Press the HOLD/ENTER key to confirm selection. 
7.  Press the CAL/MEAS key twice to return to 
measurement mode.  
NOTE: Parameter P9.1 will only be in the Set Up menu if 
you enter from the conductivity measurement mode. If 
you enter from the TDS measurement mode, Parameter 
P9.1 will not be displayed.  
 
Asc
SETUP
 P 9.0
ACA
SETUP
 P 9.1
ACA
SETUP
 YES
ACA
SETUP
MI
MR
HOLD
ENTER
HOLD
ENTER
Figure 44: P9.1-Setting 
Automatic or Manual 
Calibration