Insulet Corporation 017 Manual Do Utilizador
151
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure
that is used in such an environment.
Emissions
that is used in such an environment.
Emissions
Compliance according to
Electromagnetic environment
RF Emissions (CISPR 11)
Group 2
The Pod and the PDM emit low level electromagnetic energy (RF) in
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
CISPR B
Emissions Classification
Emissions Classification
Class B
The System is suitable for use in all establishments including domestic
establishments.
establishments.
Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. You should observe these requirements in the
use of the System.
use of the System.
Immunity against
IEC 60601-1-2
test level
test level
Compliance level
(of this device)
(of this device)
Electromagnetic environment
ElectroStatic Discharge,
ESD (IEC 61000-4-2)
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.
static discharges.
Power Frequency mag-
netic fields 50/60 Hz
(IEC 61000-4-8)
netic fields 50/60 Hz
(IEC 61000-4-8)
3 A/m
400 A/m
Suitable for most environments. Magnetic field strengths in
excess of 400 A/m would be unlikely except in close proximity
to industrial magnetic devices.
excess of 400 A/m would be unlikely except in close proximity
to industrial magnetic devices.