K-jump Health Co. Ltd. TEMPONITOR Manual Do Utilizador
11
8
9
10
Reading the Memories
Maintenance
Error Message
Specifications
Power Source
6V ; 1.5Vdc AAA size battery X4
6V ; 1.5Vdc AAA size battery X4
DC
Measuring range
90.0 F ~ 109.9 F,(32.0 C ~ 43.9 C)
90.0 F ~ 109.9 F,(32.0 C ~ 43.9 C)
Resolution
0.1 F (
0.1 C)
Accuracy
0.2°F (±0.1°C): 93.2~107.6°F (32.0 ~
42.0°C).
0.4°F (±0.2°C): other range.
Operation Frequency
13.56MHz
13.56MHz
Operation Range
4cm diameter circle with distance within
3cm.
4cm diameter circle with distance within
3cm.
Memory
24 memories
24 memories
Low battery
Replace the batteries if the low battery
indicator appears.
Replace the batteries if the low battery
indicator appears.
Power OFF
Automatically after 1 minutes of non use.
Automatically after 1 minutes of non use.
Operating environment
+60.8 F~+104.0 F(
+60.8 F~+104.0 F(
16.0 C ~
40.0 C),at
95%RH
Storage environment
-68.0 F~+122.0 F(
-68.0 F~+122.0 F(
20.0 C ~
50.0 C),at
95%RH
or
No display
No display
Error Marks
Correction
The temperature
measured was higher
than the specified
operating range.
Please measure
again.
measured was higher
than the specified
operating range.
Please measure
again.
The temperature
measured was lower
than the specified
operating range.
Please measure
again.
measured was lower
than the specified
operating range.
Please measure
again.
Low Battery Indication
Replace four 1.5V
size AAA batteries
size AAA batteries
1. The KD-2100 provides
24-set memories
including date/time
information for your
tracking of the
measurement by simply
clicking the "M" key.
including date/time
information for your
tracking of the
measurement by simply
clicking the "M" key.
2. The first click will trigger to show the
nearest measurement being taken. To
navigate previous ones, further click "M"
key to show the memory data.
navigate previous ones, further click "M"
key to show the memory data.
Note:
The memory data will be re-freshed once
the batteries are replaced. Be sure to
record the data to your note if it is
important.
record the data to your note if it is
important.
The maximum memories to be stored are
24-set, in case of the data bank is full,
the earliest one will be erased
automatically.
the earliest one will be erased
automatically.
M key
MEMORY
TIME
MEMORY
TIME
MEMORY
DATE
(M/D)
MEMORY
DATE
(M/D)
Using the Flash Light
The KD-2100equipped with
flash light which allows user
to operate during dark area.
By pressing the "Flash Light"
button, one can easily shine
the limited target area
without disturbing the
subject.
The KD-2100equipped with
flash light which allows user
to operate during dark area.
By pressing the "Flash Light"
button, one can easily shine
the limited target area
without disturbing the
subject.
Flash Light Button
Turn Off the Device
After use, you can turn off the power by
clicking the O/Ikey, or, the device will turn
off automatically after 60 seconds without
any operation.
After use, you can turn off the power by
clicking the O/Ikey, or, the device will turn
off automatically after 60 seconds without
any operation.
Do not modify or
disassemble the
unit or patch.
disassemble the
unit or patch.
Do not twist the
sensor.
sensor.
Use a cloth
moistened with
water or neutral
detergent to
clean the body
of the unit and
then wipe it dry.
moistened with
water or neutral
detergent to
clean the body
of the unit and
then wipe it dry.
Avoid thinner,
benzene, and
other harsh
cleaners.
benzene, and
other harsh
cleaners.
Avoid storage in
hightemperature,
direct sunlight, high
moisture, and dust.
hightemperature,
direct sunlight, high
moisture, and dust.
Remove the
batteries if the
unit will not be
used for a long
time.
batteries if the
unit will not be
used for a long
time.
Do not drop the
unit.
unit.
37.0
15:38
TIME
37.0
15:38
TIME
37.0
15:38
TIM
E
37.0
15:38
TIM
E
System Failure
The device is either
out of function or
unable to store
memory data. Check
with the dealer.
out of function or
unable to store
memory data. Check
with the dealer.
Transmission Failure
1. Re-position the
reader to closer
distance.
distance.
2. Replace a new patch
due to the one is
malfunction.
malfunction.
3. Replace the batteries.