Setting the Device
sensor
Adhesive tape
About the sensor
The sensors is constructed  by two parts: 
transmitter and temperature sensor.  The
transmitter is used for inducing power as 
well as data transmission; the temperature
sensor is used for body temperature
detecting.
Applying the Sensor
For fixing the sensor, a dressing (or surgical)
adhesive tape will be needed. A high quality 
tape without side-effect is recommended 
since long period (usually 24~48 hr)
adhesion.
1. Apply the sensor to ear canal, armpit or 
abdominal area below navel(depended 
on different types).
2. For armpit and body skin, fixing the 
sensor by medical adhesive tape(s).  Be
sure the firmness.
3. Subject is allowed to dress cloth rather 
than naked.
4. For infant or baby, the patch under diaper 
will benefit the measurement result since
warmer measurement environment.
sensor
Adhesive
 tape
Preparation before Measurement
Period for Heat Equilibrium 
After applying  the sensor, a 5~10 minutes 
waiting for temperature equilibrium 
between the sensor and your skin surface 
will be necessary before taking the first 
measurement.
Room
temperature
Skin surface 
temperature
5~10 min.
Start data reading
Setting °C or °F
The KD 2101 provides user to select the unit
in either °C or °F by following steps:
1. Click the "O/I" key to turn on the device. 
2. The LCD will show all 
segments first and
enter to default unit 
mode; now, you have 
5-second period for
the setting; keep
pressing the O/I key 
and the unit will 
switch to either °C or 
°F you want.
Setting Date and Time 
After installation/replacement of the 
batteries, the date and time are re-freshed 
to default value; a setting is necessary by 
following steps:
1. Under power-off mode, 
keep pressing the "M" 
key until it shows a 
flashing "2003" which is 
the setting mode of year.
2. Every clicking of the "M"
key to enter the next 
mode said month
  date 
 hour 
 minute to the
end.
3. Every clicking of the of
"O/I" key will add one to
the value for your setting. 
4. After the setting, the
device is always back to 
"power-off" mode.
M key
DATE
(M/D)
DATE
(M/D)
TIME
TIME
O/I key
(+ 1)
O/I key
(+ 1)
(Keep pressing
 O/I key for 2 sec)
TIME
TIME
Important Information
Start Measurement
Room Temperature Affects Skin
Temperature
Skin is major organ for maintenance of body 
temperature; either hot or cold environment
will cause skin temperature to rise or to fall.
It is recommended that subject is kept in a
comfortable room temperature (usually 
22~28°C (71.6~82.4 °F) for taking the
measurement.
Notice
When using the skin sensor or armpit 
sensor, wipe body skin with a dry towel . 
1. Turn on the device, the device will
perform system checking automatically 
until "ready for detecting" as figure:
2. Place the reader on the upright  direction 
of the transmitter around  3~4 cm (1,5 
inches) distance.
3. Click the "start key", the device starts 
scanning the patch temperature with 
flashing "---".
4. The result is then shown on the LCD with 
2 beep sounds.  Measurement is 
completed.
5. For another reading, just repeat the steps 
mentioned above.
with
in
3 cm
(1.5
inch)
TIME
TIME
TIME
TIME
TIME
TIME
Start
key
Insert the ear
tape patch
into ear
canal.