ComSonics Inc. 2007VMD Manual Do Utilizador

Página de 20
Qualifier X-Ray 
4
 
1 - 8 0 0 - 3 3 6 - 9 6 8 1
 
Display Screen 
The Display screen provides information on the 
battery condition, units of measure, shielding 
integrity value, and if enabled, the PASS/FAIL 
status of the measurement. A back light illumi-
nates the display. 
 
Battery Level 
A fresh set of batteries is indicated by four bars. 
Each bar represents approximately 25% runtime 
capacity. A low battery condition is indicated by 
no bars showing. The HHD automatically powers off when the battery state is low. 
Note: The accuracy of the battery level indicator is relative and will vary with the 
battery type and manufacturer. 
 
Shielding Integrity Measurement Value 
When a measurement is made, by a single press of the Button, the level of the 
ingress signal is displayed as a shielding integrity value in dB units-of-measure. 
The value range is 80 dB (very good shielding) to 25 dB (very poor shielding). 
 
PASS / FAIL Function 
The HHD can be setup with a measurement threshold value to determine if the 
integrity reading is acceptable. PASS is displayed if the measured integrity value 
is higher (better) than the threshold value. FAIL is displayed if the measured integ-
rity value is lower (worse) than the threshold value. The PASS/FAIL threshold 
value is user selectable. This function can be disabled as described on the next 
page. 
 
     
 
 
Threshold Value Setup 
To access the threshold value setup mode, press the Button twice within one 
second when the HHD is on. The PASSFAIL readout activates and two digits 
display the current threshold value. Press and hold the Button to scroll to the de-
sired threshold value. Release the Button to stop rapid scrolling and then use 
single presses of the Button to the desired value. If a lesser value is desired, con-
tinue to hold the Button as the value wraps around from 80 and then back down to 
25 dB. Release the Button and allow the HHD to auto power off (six seconds). At 
the next power up, the newly set threshold value is enabled and the PASS or the 
FAIL indicator will display accordingly. See the next sections for the procedure to 
disable or enable the PASS / FAIL function.