Agfa Gevaert N V 5364A Manual Do Utilizador

Página de 266
112
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
Film calibration
The film calibration combines Dmax and sensitometry calibration, and is
activated as one single procedure. Separate Dmax and sensitometry
calibration steps are not required.
For more general information on maintaining image quality, refer to
 on page
.
To start the film calibration:
1
Perform step 1 to 2 of the 
 procedure on page
 
to select the ‘Calibration’ screen.
2
Press the Confirm key to select ‘Film’.
The ‘Select input tray’ screen appears:
3
Press the Up/Down arrow keys to select an input tray, followed by the Confirm 
key (e.g. choose upper input tray).
The ‘Film calibration in progress’ screen appears:
The counter indicates the progress of the calibration process. The calibration film is 
measured automatically by the built-in MDM (Macro DensitoMeter). The 
Drystar 5500 is being calibrated during the printing process.
In the procedure, we will assume a film calibration on the upper input tray. The 
procedure for the lower input tray is identical.