Agfa Gevaert N V 5364A Manual Do Utilizador

Página de 266
113
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
4
In case the film calibration has failed, the following message will appear.
Press the Confirm key to retry the calibration.
5
When the film calibration has completed succesfully, this indicates that the 
calibration procedure has been executed succesfully. This does not always mean 
that the result of the calibration is succesful.
Three cases are possible:
Calibration successful, results OK.
The obtained Dmax (x.xx) after calibration is displayed:
Calibration successful, but Dmax differs more than 2% from the target 
value.
The obtained Dmax (x.xx) after calibration is displayed:
Calibration procedure succesful, but limited by film-Dmax. The obtained 
Dmax does not meet  the required system Dmax and is limited by the film 
sensitometry.
The obtained Dmax (x.xx) after calibration is displayed:
6
Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the 
calibration procedure again. Otherwise, the old calibration parameters 
will be used.
Blinking
Blinking
X.XX
Blinking
Default Densitometer 
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer 
e.g. Macbeth TR924
X.XX
Blinking
Default Densitometer 
e.g. Macbeth TR924