PAIN THERAPY 
 
 
 
NOTE:  
Follow the Fitzpatrick Skin Type Scale noted below when performing pain 
 
             therapy procedures.  Select the appropriate clinical parameter and reduce 
 
             power as necessary for patient comfort.
 
 
Fitzpatrick Skin Type Scale  
 
TYPE I
 
Highly sensitive, always burns, never tans.  Example:  Red hair with 
freckles
 
 
TYPE II
 
Very sun-sensitive, burns easily, tans minimally.  Example:  Fair-
skinned, fair-haired Caucasians
 
 
TYPE III
 
Sun-sensitive skin, sometimes burns, slowly tans to light brown.  
Example:  Darker Caucasians
 
 
TYPE IV
 
Minimally sun-sensitive, burns minimally, always tans to moderate 
brown.  Example:  Mediterranian-type Caucasians
 
 
TYPE V
 
Sun-insensitive skin, rarely burns, tans well. Example:  Some Hispanics, 
some Blacks
 
 
TYPE VI
 
Sun-insensitive, never burns, deeply pigmented.  Example:  Darker 
Blacks
 
 
Using the Whitening/Contour Handpiece
:  Recommended Clinical Settings 
  
The EPIC 10 diode laser, in conjunction with the Whitening/Contour Handpiece, is designed to 
provide near-infrared laser energy to tissue for the purpose of elevating tissue temperature 
and providing for temporary relief of pain conditions as stated in the Indications for Use. 
 
Affected muscles and/or joints have to be exposed to an adequate level of therapeutic energy 
over a short period of time to be effective.  Two main therapeutic power settings are 
recommended for these treatments: 
 
 
►   2.75W CW:  place the handpiece in contact with the skin and apply laser energy for 10 
minutes continuously. 
 
►   5.50W CW:  place the handpiece approximately 2 – 3mm away from the skin surface 
(non-contact).  Exposure time remains the same at 10 minutes continuously. 
 
Patients should be monitored for discomfort.  If discomfort is reported at any time during the 
treatment, there are several options: 
 
a)   Decrease the power setting to 2.75W; or 
 
b)   Defocus the energy by moving the handpiece further away from the sking; or 
 
c)   Stop treatment.