Using the Deep Tissue Handpiece
: Recommended Clinical Settings
TYPE II
Set Power to 4.0W CW for 2 minutes, followed by Power at 3.25W for 3
minutes and then set Power to 2.75W for the remaining 5 minutes
minutes and then set Power to 2.75W for the remaining 5 minutes
TYPE III
Set Power to 4.5W CW for 2 minutes, reduce Power to 4.0W for the
remaining 8 minutes
remaining 8 minutes
TYPE IV
Set Power to 4.0W CW for 2 minutes, reduce Power to 3.5W for the
remaining 8 minutes
remaining 8 minutes
TYPE V
Set Power to 3.5W CW for 2 minutes, reduce Power to 3.0W for the
remaining 8 minutes.
remaining 8 minutes.
TYPE I
Less common, settings are not available
TYPE VI
Less common, settings are not available
NOTE:
Some patients may require more than one laser application or a series of
treatments before significant improvement is reported. Repeat this therapy as
necessary and monitor progress of the patient’s condition throughout the
the treatment.