Biolase Inc. EPIC Manual Do Utilizador

Página de 45
35 | 
P a g e
                                                                             
5 4 0 0 3 2 1 R e v .   X :   D R A F T
 
 
Using the Deep Tissue Handpiece
:  Recommended Clinical Settings 
 
 
TYPE II
 
Set Power to 4.0W CW for 2 minutes, followed by Power at 3.25W for 3 
minutes and then set Power to 2.75W for the remaining 5 minutes
 
TYPE III
 
Set Power to 4.5W CW for 2 minutes, reduce Power to 4.0W for the 
remaining 8 minutes
 
TYPE IV
 
Set Power to 4.0W CW for 2 minutes, reduce Power to 3.5W for the 
remaining 8 minutes
 
TYPE V
 
Set Power to 3.5W CW for 2 minutes, reduce Power to 3.0W for the 
remaining 8 minutes.
 
TYPE I
 
Less common, settings are not available 
 
TYPE VI
 
Less common, settings are not available
 
 
 
NOTE:
  Some patients may require more than one laser application or a series of 
 
              treatments before significant improvement is reported.  Repeat this therapy as
 
              necessary and monitor progress of the patient’s condition throughout the
 
              the treatment.