Introduction 
The ActiGraph Link activity monitors provide objective measurements of human activity and are used in 
many research and clinical applications.  They include a microelectromechanical system (MEMS) based 
accelerometer, gyroscope, magnetometer, and a capacitive touch based wear sensor as data collection 
end points.  No formal training is required to use the ActiGraph Link and patients may be the intended 
operator.  This document will serve as the only necessary training guide needed to get full functionality 
from the device.  
Specifications and Functionality of the ActiGraph Link 
Data 
The  Link includes two accelerometers, gyroscope, and magnetometer sensors.  Data  will  not  be 
negatively affected by lint, dust, or direct sunlight.  
Acceleration 
Acceleration is measured by two accelerometers; a primary that is utilized for all historical and standard 
data collection efforts and a secondary which is used in very specific use cases where a higher dynamic 
range is necessary. 
Parameter 
Primary 
Secondary  Units 
Axis 
 
Sensitivity 
0.5 
mg/LSB 
Dynamic Range 
+/-16 
+/- 8 
Table 1 – Accelerometer Key Specifications 
Acceleration data from the primary sensor is sampled at rates ranging from 30 Hz to 100 Hz (user 
selectable) and the secondary accelerometer data is sampled at a fixed 100 Hz.  Output from both 
devices are stored in a raw, non-filtered/accumulated format in the units of gravity (g’s).  This data is 
stored directly into a non-volatile flash memory.  The accelerometer is very sensitive and is accurate 
within +/- 0.5% of the data collected.  
Gyroscope 
A three axis gyroscope collects angular velocity in units of degrees/sec.  Data is sampled at 100 Hz. 
Parameter 
Value 
Units 
Axis 
 
Sensitivity 
0.06 