Electromagnetic Compatibility
The performance of this device under electromagnetic environment complies with
IEC60601-1-2 (2007).
IEC60601-1-2 (2007).
Precautions for Safe Operation under Electromagnetic Influence
CAUTION
If any sorts of electromagnetic wave, magnetic field, or static electricity
exist around the equipment, noise interference or malfunction of the
device may occur. If any unintended malfunction or noise occurs in
monitoring, check the magnetic influence and take appropriate
countermeasures
The following are examples of the common cause and
countermeasures.
z
exist around the equipment, noise interference or malfunction of the
device may occur. If any unintended malfunction or noise occurs in
monitoring, check the magnetic influence and take appropriate
countermeasures
The following are examples of the common cause and
countermeasures.
z
Cellular Phone
The radio wave may cause malfunction to the device.
Cellular phones and radio sets should be turned off in the room
(building) where a medical device is located.
Cellular phones and radio sets should be turned off in the room
(building) where a medical device is located.
z
Static Electricity
In a dry environment (room), static electricity is likely to occur. Take the
following countermeasures.
following countermeasures.
y Both operator and patient should remove any static electricity before
entering the room.
y Humidify the room.