Electromagnetic Compatibility 
 
The performance of this device under electromagnetic environment complies with 
IEC60601-1-2 (2007). 
 
Precautions for Safe Operation under Electromagnetic Influence 
CAUTION 
If any sorts of electromagnetic wave, magnetic field, or static electricity 
exist around the equipment, noise interference or malfunction of the 
device may occur. If any unintended malfunction or noise occurs in 
monitoring, check the magnetic influence and take appropriate 
countermeasures 
The following are examples of the common cause and 
countermeasures. 
Cellular Phone 
The radio wave may cause malfunction to the device.   
Cellular phones and radio sets should be turned off in the room 
(building) where a medical device is located. 
Static Electricity 
In a dry environment (room), static electricity is likely to occur. Take the 
following countermeasures. 
y  Both operator and patient should remove any static electricity before 
entering the room. 
y  Humidify the room.