Fukuda Denshi Co Ltd HLX801 Manual Do Utilizador

Página de 40
xi 
Compliance to the Electromagnetic Immunity (1) 
The HLX-801 is intended for use in the electromagnetic environment specified in 
below tables. It should be assured that the HLX-801 is used in such an environment. 
 
Immunity Test
IEC60601-1-2 
Test Level 
Compliance Level
Electromagnetic 
Environment - 
Guidance 
Electrostatic 
Discharge  
(ESD) 
IEC61000-4-2 
±
6kV  contact 
±
8kV  air 
±
6kV  contact 
±
8kV  air 
Floors should be wood, 
concrete or ceramic tile. 
If floors are covered 
with synthetic material, 
the relative humidity 
should be at least 30%.
Electrical fast 
transient/burst 
IEC61000-4-4 
±
2kV 
  Power supply 
lines 
±
1kV 
  for input/output 
lines 
±
2kV 
  Complies 
±
1kV 
  for input/output 
lines 
Mains power quality 
should be that of a 
typical commercial or 
hospital environment. 
Power Frequency 
(50/60Hz) 
Magnetic Field 
IEC61000-4-8 
3A/m 3A/m 
Power 
frequency 
magnetic fields should 
be at levels 
characteristic of a 
typical location in a 
typical commercial or 
hospital environment. 
Note: U
is the AC mains voltage prior to application of the test level.