Compliance to the Electromagnetic Immunity (1)
The HLX-801 is intended for use in the electromagnetic environment specified in
below tables. It should be assured that the HLX-801 is used in such an environment.
below tables. It should be assured that the HLX-801 is used in such an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic
Environment -
Guidance
Electrostatic
Discharge
(ESD)
IEC61000-4-2
Discharge
(ESD)
IEC61000-4-2
±
6kV contact
±
8kV air
±
6kV contact
±
8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
transient/burst
IEC61000-4-4
±
2kV
Power supply
lines
lines
±
1kV
for input/output
lines
lines
±
2kV
Complies
±
1kV
for input/output
lines
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
should be that of a
typical commercial or
hospital environment.
Power Frequency
(50/60Hz)
Magnetic Field
IEC61000-4-8
(50/60Hz)
Magnetic Field
IEC61000-4-8
3A/m 3A/m
Power
frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.