Panorama™ Operating Instructions
0070-10-0690-01
6 - 15
Patient Alarm Functions
Alarm Responses Tab (Patient)
5. The Save To Event button automatically saves alarm data in the Events database 
when an alarm condition occurs.
• The ON option automatically saves alarm data during an alarm condition.
• The OFF option does not save alarm data during an alarm condition.
6.3.3
Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings 
Select the Restore Previous Settings sidebar button to return patient display settings to 
those in use when the tab was opened.
NOTE:
The Restore Previous Settings sidebar button is not available 
at the ViewStation.
Suspend All Alarms (Optional)
Select the Suspend All Alarms sidebar button to suspend all numeric physiological, lethal 
arrhythmia, non-lethal arrhythmia, and technical alarms at the Panorama Central Station for 
the selected patient tile. The Suspend All Alarms sidebar button is only shown if the 
Suspend Alarms option is ON in the System Alarms tab.
NOTE:
The Suspend All Alarms sidebar button is not available at 
the ViewStation or at the WorkStation.
When the Suspend All Alarms sidebar button is selected:
• The Suspend All Alarms sidebar button changes to white with black text.
• The All Alarms Suspended message displays in red text in the Patient Status line of 
the patient tile. For additional information regarding the Patient Status line, refer to the 
‘‘Patient Status Line’’ on page 2-10.
To resume the monitoring of alarms, select the Suspend All Alarms sidebar button.
• When the Suspend All Alarms sidebar button is disabled, the button changes to blue 
with white text. Alarms are now re-enabled for the selected patient.
Suspend Non-Lethal 
Select the Suspend Non-Lethal sidebar button to suspend all numeric, technical and 
non-lethal arrhythmia alarms for the selected patient tile. Lethal arrhythmia alarms continue to 
be monitored at the Panorama Central Station when non-lethal arrhythmia alarms are 
suspended.