Alarm Responses Tab (Patient)
Patient Alarm Functions
6 - 16
0070-10-0690-01
Panorama™ Operating Instructions
NOTE:
The Suspend Non-lethal sidebar button is not available at 
the ViewStation or at the WorkStation.
NOTE:
The Brady alarm will not be suspended when the Suspend 
Non-Lethals sidebar button is selected.
When the Suspend Non-Lethal sidebar button is selected:
• The Suspend Non-Lethal sidebar button changes to white with black text.
• The Non-Lethal Alarms Suspended message displays in yellow text in the Patient 
Status line of the patient tile. For additional information regarding the Patient Status line, 
refer to the ‘‘Patient Status Line’’ on page 2-10.
To resume the monitoring of non-lethal alarms, select the Suspend Non-Lethal sidebar 
button.
• When the Suspend Non-Lethal sidebar button is disabled, the button changes to blue 
with white text. Non-lethal alarms are now re-enabled for the selected patient.
Alarm Setup (Optional)
Select the Alarm Setup sidebar button to configure the ST Alarm Delay and the Alarm 
Delay, for the selected patient tile, when using a Telepack device. 
NOTE:
The Alarm Setup sidebar button is not available at the 
ViewStation.
To configure alarm delay:
1. Select the Alarm Setup button. The Alarm Setup dialog box (shown in FIGURE 6-8) 
is displayed. 
 
FIGURE 6-8 Alarm Setup Dialog Box 
2. Select the ST Alarm Delay time period for the selected patient’s alarm. ST Alarm 
Delay is the period of time that the Panorama Central Station delays the notification of 
an ST alarm. ST Alarm Delay options include 30 seconds, 45 seconds, 1 minute, 90 
seconds, 2 minutes and 3 minutes.
3. Select the Alarm Delay time period for the selected patient’s alarm. Alarm Delay is the 
period of time that the Panorama Central Station delays the notification of an alarm. 
Alarm Delay options include None, 1 second, 2 seconds, 3 seconds, 4 seconds, 5 
seconds, 6 seconds, 7 seconds and 8 seconds.
NOTE:
The Alarm Delay setting does not affect the alarm 
notification for ST, Apnea, or Arrhythmia.
NOTE:
Changing the alarm delay becomes effective immediately 
and affects the patient’s current alarm delay setting.