Multi-Tech Systems 92U05E06800 Manual Do Utilizador

Página de 84
Responding to alerts 
Add gel 
 
After you receive a treatment to correct an abnormal 
rhythm, you may see this message telling you to add gel. 
Follow the instructions below to add gel. 
While you’re wearing the LifeVest, tap the help button 
 
for reminders about what to do. 
Tap OK to resume normal monitoring. 
To add gel 
1    Remove the battery from the monitor, then remove the electrode belt and 
garment from your body. 
2    Locate the packets of gel you got with the LifeVest system. Add one-half 
packet of gel to each therapy pad, directly onto the mesh material of each 
pocket. Apply to the rear pads and the front pad (but not to the ECG 
electrodes.) 
 
3    Put on the electrode belt and garment. Refer to section 4. 
4    Put the battery into the monitor and follow the normal startup routine. 
1-800-543-3267 