LifeVest 4000 Patient Manual
Replace belt
After you receive a treatment to correct an abnormal
rhythm, you may see this message telling you to replace
the belt.
rhythm, you may see this message telling you to replace
the belt.
Continue to wear the belt until you get the replacement belt.
The belt is still functional and can provide additional
treatment if needed.
The belt is still functional and can provide additional
treatment if needed.
When you receive the new belt, follow the instructions
below to replace the belt.
below to replace the belt.
While you’re wearing the LifeVest, tap the help button
for reminders about what to do.
Tap OK to resume normal monitoring.
To replace the electrode belt
1 Remove the battery from the monitor.
2 Disconnect the electrode belt from the monitor.
3 Remove the electrode belt and garment from your body.
4 Disassemble the electrode belt from the garment, assemble the new
electrode into a clean garment, and put on the assembled electrode belt and
garment. Refer to section 4.
garment. Refer to section 4.
5 Connect the electrode belt to the monitor.
6 Put the battery into the monitor and follow the normal startup routine.
Page 5-16