LifeVest 4000 Patient Manual 
Replace belt 
 
After you receive a treatment to correct an abnormal 
rhythm, you may see this message telling you to replace 
the belt. 
Continue to wear the belt until you get the replacement belt. 
The belt is still functional and can provide additional 
treatment if needed. 
When you receive the new belt, follow the instructions 
below to replace the belt. 
While you’re wearing the LifeVest, tap the help button 
 
for reminders about what to do. 
Tap OK to resume normal monitoring. 
To replace the electrode belt 
1    Remove the battery from the monitor. 
2    Disconnect the electrode belt from the monitor. 
3    Remove the electrode belt and garment from your body. 
4    Disassemble the electrode belt from the garment, assemble the new 
electrode into a clean garment, and put on the assembled electrode belt and 
garment. Refer to section 4. 
5    Connect the electrode belt to the monitor. 
6    Put the battery into the monitor and follow the normal startup routine. 
Page 5-16 